LYRA x2

Top notch Performance of Non-Invasive and Invasive Ventilation for all Patient Categories

LYRA x2 is a premium non-invasive turbine driven ventilator with no compromise on the performance in invasive ventilation.

LYRA x2

Top notch Performance of Non-Invasive and Invasive Ventilation for all Patient Categories

LYRA x2 is a premium non-invasive turbine driven ventilator with no compromise on the performance in invasive ventilation.

User can easily switch between NIV- and IV-modes by UI operation only.
Comprehensive parameter monitoring describes the full scenario of patient’s status to the care giver.

In a busy ICU it is imperative to give the desired mechanical ventilation to the patient.
An 18.5 inch vertical layout touchscreen display makes operating of the ventilator smooth & easy.

Power Battery Backup External AC power supply Input voltage: 100-240 V

Input frequency: 50/60 Hz

Input current: 2.5 A Max

Build-in Lithium-ion battery, 3 hours (Powered by one new fully-charged battery in standard working condition)

Communication interface: Rs232, Ethernet, USB port, Nurse call, CO2 calibration connector

Gas supply: O2 Pipe Connector: DISS Gas supply pressure:

Dimensions: 327 mm x 310 mm x 493 mm

Weight: Approximately 21 kg

Display Size: 18.3 Color active matrix TFT touch

Display Resolution 1080 x 1980 pixels

Brightness: Adjustable

Ventilation Specifications

Patient Type: Adult, Pediatric Invasive Ventilation Mode:

VCV (Volume Control Ventilation)

PCV (Pressure Control Ventilation)

VSIMV (Volume Synchronized Intermittent Mandatory Ventilation)

PSIMV (Pressure Synchronized Intermittent Mandatory Ventilation)

CPAP/PSV (Continuous Positive Airway Pressure/Pressure Support Ventilation)

PRVC (Pressure Regulated Volume Control) V + SIMV (PRVC + SIMV)

BPAP (Bilevel Positive Airway Pressure)

APRV (Airway Pressure Release Ventilation) Apnea Ventilation Non-invasive Ventilation Mode

CPAP (Continuous Positive Airway Pressure)

PPS (Proportional Pressure Support)

S/T (Spontaneous and Timed)

VS (Volume Support, similar like AVAPS )

Controlled Parameters

O2 %: 21-100% (increments of 1%)

VT (Tidal Volume): Adult: 100-2000 mL (increments of 10 mL) / Pediatric: 20-300 mL (increments of 1 mL)

f (Ventilation frequency): 1-80 bpm (increments of 1 bpm)

fSIMV (Ventilation frequency in SIMV mode): 1-80 bpm (increments of 1 bpm)

I:E range: 4:1-1:10 (increments of 0.5)

Tinsp (Inspiratory time): 0.20-10 s (increments of 0.05 s)

Tslope (Time of Pressure Rising): Thigh 0-2.00 s (increments of 0.05 s) Tlow: 0.2-30 s (increments of 0.1 s) Tpause: 0.2-30 s (increments of 0.1 s)

ΔPinsp: 5%-60% (increments of 1%), Off

ΔPsupp: 0-60 cm H2 O (increments of 1 cm H2 O) Phigh: 0-60 cm H2 O (increments of 1 cm H2 O) Plow: 0-45 cm H2 O (increments of 1 cm H2 O)

PEEP: 1-45 cm H2 O (increments of 1 cm H2 O), Off

Flow trigger: 0.5-15 L/min (increments of 0.1 L/min)

Pressure trigger: -10 to -0.5 cm H2 O (increments of 0.5 cm H2 O)

Auto CPAP: 4-25 cm H2 O (increments of 1 cm H2 O)

EPAP: 4-25 cm H2 O (increments of 1 cm H2 O)

IPAP: 4-20 cm H2 O (increments of 1 cm H2 O)

Max P (VS maximum IPAP): 6-40 cm H2 O (increments of 1 cm H2 O)

Max P (PPV maximum pressure limit): 5-40 cm H2 O (increments of 1 cm H2 O)

Max V (PPV maximum volume limit): 200-3500 mL (increments of 5 ml)

Max E: 0-100 cm H2 O/L (increments of 1 cm H2 O/L) Max R: 0-50 cm H2 O/L (increments of 1 cm H2 O/L)

Apnea Ventilation

Vtvapnea: Adult: 100-2000mL (increments of 10 mL) / Pediatric: 20-300 mL (increments of 1 mL) ΔPapnea: 5-60 cm H2 O (increments of 1 cm H2 O)

Fapnea: 1-80 bpm (increments of 1 bpm)

Apnea Tinsp: 0.20-10 s (increments of 0.05 s)

Monitored parameters

Paw

Oxygen concentration

WOB

Ppeak

VTe

spn

RSBI

Pplat

VTe/IBW

NIF

Pmean

PEEP

Ftotal

PEEPi

Insp Flow

PIP

Exp Flow

EPAP

Tot.leak

Continuous Flow (O2 Therapy)

Alarm settings

Tidal Volume: High: Adult: 110-4000 mL, Off / Pediatric: 25-600 mL,Off Low: Adult: 50-4000 ml, Off / Pediatric: 10-600 mL,Off

Minute Volume: High: Adult: 0.2-100.0 L/min / Pediatric: 0.2-60.0 L/min Low: Adult: 0.1-50.0 L/min / Pediatric: 0.1-30.0 L/min

Airway pressure: High: 10-85 cm H2 O Frequency: High: 1-150 bpm, Off Inspired oxygen (FiO2 ): High: Actual reading +8% (High-pressure oxygen) / 2%-100% (Low-pressure oxygen)

Apnea alarm time: Low: Actual reading -8% (High-pressure oxygen) / 18%-98% (Low-pressure oxygen) / 5-60 s